A CCD is the mainstream imaging component of digital PCR at present, with mature technology and better imaging quality than CMOS; CMOS is a low-end component with relatively low cost, poor imaging quality, and signal crosstalk between fluorescence channels.
-
A The sample throughput can provide three models of 8, 16, and 32 samples, and the detection channel can provide three models of three, four, and five channels, which can be freely matched according to needs, and can be compatible and upgraded all-in-one machine.
-
A In theory, the agreement rate can reach 100% without considering other factors. The actual situation needs to consider the differences between methodologies, as well as the entire detection system, human operation, and various factors of error.
-
A For users who need to calibrate the equipment once a year, our company can provide corresponding services and provide calibration reports, and the relevant fees can be negotiated with the corresponding business personnel.
-
A We can arrange technicians to implement the performance of instruments and equipment, and provide corresponding verification plans and verification reports. For the project performance verification that the clinical laboratory needs to do, our company can arrange technical personnel to assist in the whole process.
-
A For ventilation, the air is filtered while pressurizing to generate water-in-oil droplets.
-
A 1. The operation time is too long after the droplet generation oil is added ( recommended within 10 minutes); 2. Try to avoid the bubbles introduced by the sample addition operation ; 3. The accuracy of the sample addition; 4. The interference of impurities in the sealing cover; 5. Ambient temperature and humidity.
-
A The temperature rise and fall of the TC equipment is completed by Peltier , and the temperature is precisely controlled by a metal temperature control probe (one probe under each chip of 16 fluxes).